Загрузка страницы

Isoclean® Healthcare Platform Isolator | HPI-G3 Glove Leak Test | Esco Pharma

Watch the full video to learn how to conduct HPI-Gen 3 Glove Leak Test.

1. Turn on the Main Compressed Air Line
2. Install the Glove Leak Test Disc
3. Turn on the Pressure Valve
4. Once the disc is already secured, turn off the Pressure valve and remove the Pneumatic Hose.
5. Open the Supply Pressure Valve to inject air up to 250 Pa.
6. Upon reaching 250 Pa, close the Pressure valve and remove the Pressure Tube.
7. Record the pressure decay every 10 minutes for 30 minutes.
8. Once done, remove the Glove Leak Test Disc.

------------------------------------------------------------------

The Esco Lifesciences Group, headquartered in Singapore, is building a synergistic ecosystem of life science tools, diagnostics, therapeutics, and technologies to help improve human lives.

Visit our new and improved website: http://www.escolifesciences.com

Subscribe to our channel: http://bit.ly/EscoLifeSciences 

Follow us on our social media accounts to get the latest news and updates. 

Facebook: https://www.facebook.com/escolifesciences
Instagram: https://www.instagram.com/escolifesciences
Twitter: https://twitter.com/EscoLifesci
LinkedIn: https://www.linkedin.com/company/esco-micro-pte-ltd
TikTok: https://www.tiktok.com/@escolifesciences

#EscoLifesciencesGroup