Загрузка страницы

Calibration and Configuration of Epsilon HT668 For Error Show Temperature High and Low

If it Shows Error Like Low Temperature and High Temperature Then Don’t Worry It is
Need to Configure and Caliber Before Send Return and Give Bad Review Please
Consult with Our Customer Care, They Will Guide You How to Solve This Problem.
Here are Some Step to Configure Epsilon Infrared Thermometer HT 668 Model for
Solve Problem of Show High and Low Temperature.
Step-1-Press Set and Up (↑) Button Together
Step-2-Press Trigger One Time
Step-3-There are number is blinking
Step-4-Press Trigger Again
Step-5-It is show Calibration 1 That means Set Lower
Temperature, For Setting Lower Temperature Take Gun in Front of Air
Conditioning or Freezer Make sure Freezer Door is open
Step-6-Wait for Temperature Go Below 35
Step-7-Press Trigger
Step-8-There are show Caliber 2 Option Here you need to set Higher Temperature
Step-9-Take Gun In front of high Temperature Object Make Sure Higher Temperature
Is around 38 or more
Step-10-Press Trigger Two time Then it is off Then It is Restart and Work Again

Видео Calibration and Configuration of Epsilon HT668 For Error Show Temperature High and Low канала Epsilon DigiTech
Показать
Комментарии отсутствуют
Введите заголовок:

Введите адрес ссылки:

Введите адрес видео с YouTube:

Зарегистрируйтесь или войдите с
Информация о видео
1 июля 2020 г. 15:10:25
00:01:23
Яндекс.Метрика