Загрузка страницы

Regenlab PRP - Platelet Count Test

Whole Blood PLT = 254,000 per microliter (10cc)
PRP Sample #1 = 61,000 per microliter (5cc @ 0.24X PLT Fold)
PRP Sample #2 = 100,000 per microliter (5cc @ 0.39X PLT Fold)
The video is unedited and results are shown at 3:35 and 6:21.
Product Info
Manufacturer: Regen Lab
Country of Origin: Switzerland
Product Name: Regenkit BCT
Product Type: Single Spin Separator Gel
Product Material: Glass
Product Size: 10ml
Device Classification: FDA Cleared Class 2
https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfRL/rl.cfm?lid=330886&lpcd=ORG

Website: www.regenlab.com

We tested 2 Platelet-Rich Plasma samples using Regenlab BCT kit using the exact instructions provided in the kit. Whole blood platelet count is 254,000 per microliter. Our assessment of the Regenlab system is as follows:

Strength:
- Regen BCT 10cc tubes are effective at removing WBC and RBC.
- Easy to Draw Blood
- Easy to use Centrifuge

Weakness:
- The small initial blood sample (10cc) combined with low platelet yield, lack of removal of Platelet-Poor Plasma (PPP) and ineffective recommended suspension technique makes it difficult to increase platelet concentration above whole blood/baseline platelet counts.
- Technically not making platelet-rich plasma. Actual concentration much lower than 1.6X PLT concentration being advertised.
- Some floating clotted platelets did not make it into the syringe, likely contributing to low counts.

Видео Regenlab PRP - Platelet Count Test канала PRP Truth Project